Calibration-free EPI trajectory error correction by k-space data consistency

Julianna D. Ianni1,2 and William A. Grissom1,2,3,4

1Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, United States, 2Vanderbilt University Institute of Imaging Science, Vanderbilt University, Nashville, TN, United States, 3Department of Radiology and Radiological Sciences, Vanderbilt University, Nashville, TN, United States, 4Department of Electrical Engineering, Vanderbilt University, Nashville, TN, United States

A method for automatic correction of EPI trajectory errors is presented. The method is an iterative parallel imaging reconstruction which uses k-space data consistency to correct image artifacts. An advantage of the method is that it requires no calibration data and allows correction of non-static errors such as those due to gradient coil heating.