Meeting Banner
Abstract #2602

Design of 5-channel On-coil Shimming Coil for Rat Brain MRI Guided High Intensity Focused Ultrasound: a preliminary study

Jo Lee1, Jianhong Wen1, Chao Zou1, Xiaoqing Hu1, Xiaoliang Zhang2,3, Xin Liu1, and Ye Li1

1Lauterbur Imaging Research Center, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen City, People's Republic of China, 2Department of Radiology and Biomedical Imaging, University of California San Francisco, CA, United States, 3UCSF/UC Berkeley Joint Graduate Group in Bioengineering, San Francisco, CA, United States

Signal-to-noise ratio (SNR) and homogeneity of the static magnetic field B0 are critically important in achieving accurate temperature measurement for MRI guided HIFU. In this work, a 5-channel shimming coil on the top of a 3-channel RF coil array at 3T has been fabricated to increase the B0 homogeneity and the SNR in region of interest. Phantom studies have demonstrated the the capability of the proposed shimming coil to improve B0 homogeneity and SNR of the images, which will lead to accurate temperature mapping.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords