Abstract #2804

Validation and Error Quantification of Pulmonary Artery 4D Flow-MRI in a digital broadband 3T MR setup

Thekla Helene Oechtering1, Clara Berlin1, Malte Sieren1, Daniel Droemann2, Joerg Barkhausen1, and Alex Frydrychowicz1

1Clinic for Radiology and Nuclear Medicine, University Hospital Schleswig-Holstein, Luebeck, Germany, 2Medical Clinic III: Pulmonology, University Hospital Schleswig-Holstein, Luebeck, Germany

4D Flow-MRI offers unique possibilities for the diagnosis of pulmonary hypertension. A validation of the sequence on a digital broadband 3T MR system is essential before introduction into clinical routine. This study compared 4D Flow–MRI derived stroke volume (SV) and maximal flow velocity (Vmax) in the pulmonary arteries of 23 healthy volunteers to 2D phase contrast MRI, right ventricular stroke volume (RVSV), conservation of mass-analysis (COM), and to a static phantom for additional error analysis. Results revealed clinically acceptable differences with a trend of phantom correction to improve results except for the COM-analysis.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords