SNR and volume characterization of RF coils: A simple procedure and an automatic post-processing tool for a straightforward comparison

laurent Mahieu-Williame1, Sophie Gaillard1, Denis Grenier1, Radu Bolbos2, Jean-Baptiste Langlois2, Franck Lamberton2, Sorina Camarasu-Pop1, and Olivier Beuf1

1Univ.Lyon, CREATIS, CNRS UMR 5220, Inserm U1206, INSA-Lyon, UJM- Saint Etienne Université Claude Bernard Lyon 1, Villeurbanne, France, lyon, France, Metropolitan, 2CERMEP – imagerie du vivant, Lyon, France, lyon, France, Metropolitan

To choose the most suitable RF coil available for an MRI study, we propose a procedure which uses a calibrated phantom, a 3D gradient-echo sequence, and an automatic post-processing tool available on the web. This tool generates a report which contains the measurement of a SNR with uniform volumes located in the depth. The post-processing could be done on MR images acquired on most main MRI vendors (Siemens, GE, Philips and Bruker) with prior verification of applied scaling or filtering. RF coil characterization results performed on at 4.7T and 7T were compared. The tool can be used for quality control.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords