The potential of a 256-Channel receive-only Array Coil for accelerated Cardiac Imaging at 3T

Bernhard Gruber1,2, Arjan D. Hendriks 1, Cezar B.S. Alborahal3, Bas Brussen3, Tim Leiner1, Gustav Strijkers1, Dennis W. J. Klomp1,3, and Martijn Froeling1

1Department of Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Institute of Biomedical Mechatronics, Johannes Kepler University, Linz, Austria, 3MR Coils B.V., Zaltbommel, Netherlands

High-density coil arrays can be used to accelerate MRI. Here we present the results from measurement-based extrapolations of a 256 Channel Cardiac Array Coil obtained by 16 sequential scans of a 16 Channel Array to assess acceleration performance and sensitivity constraints for 3T MRI. With element sizes of 55 mm x 33 mm, tissue load remains dominant at the 3T Larmor frequency of water, while SENSE accelerations can go up to 20-fold at low g-factors. These results motivate the design of a 256 channel cardiac array for accelerated 3T MRI.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords