Meeting Banner
Abstract #4420

A new RF coil for foot and ankle imaging at 7T MRI

Tales Roberto de Souza Santini1, Junghwan Kim1, Sossena Wood1, Narayanan Krishnamurthy1, Shailesh Raval1, and Tamer Ibrahim1

1University of Pittsburgh, Pittsburgh, PA, United States

7T Foot and ankle has been previously explored1. A four-channel Tic-Tac-Toe (TTT) transmit-only (Tx-only) array2,3 and a four-channel receive-only array were developed and combined to demonstrate the feasiability of high signal-to-noise ratio (SNR) foot and ankle imaging at 7T MRI. The experimental measurements of magnetic field distribution responsible for excitation (B1+) is in agreement with the finite-difference time-domain (FDTD) simulations and the SAR is within the FDA safety limits for the entire lower leg and foot (peak SAR ~ 7W/Kg/10g of tissue, average SAR ~ 1W/Kg/10g of tissue per continuous average field B1+ = 1.97uT). In-vivo proton density TSE and T2DESS images were acquired.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords