Abstract #0642

Validating and Measuring Transfer Functions of Straight Wires using a Combination of an Electro-optic Field Sensor and Simulation

Thomas Lottner1, Simon Reiss1, Ali Caglar Özen1,2, Michael Bock1, and Andreas Bitzer1,3

1Department of Radiology, Medical Physics, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, 2German Consortium of Translational Cancer Research Freiburg Site, German Cancer Research Center (DKFZ), Heidelberg, Germany, 3BIOLAB Technology AG, Zürich, Switzerland

To investigate deep brain stimulators, pacemakers or other elongated structures the transfer function was proposed to characterize these devices. A electro-optic E-field sensor is used to acquire 2D field data of copper wires excited at one tip for 64 and 123MHz. The data is compared to simulations and transfer functions are calculated from these simulations. These are compared to experimentally acquired transfer functions.