Catheter visualization and guidance under MRI guidance can be challenging, and conventional MR tracking coils decrease the flexibility and increase the profile of catheters. We used aerosol jet deposition to print a capacitor and an inductor with a double helix geometry on a polymer catheter. bSSFP sequences with flip angles 5°, 15°, and 90°, and a B1+ map were acquired at 3T, and CNR and amplification were measured. The marker demonstrated good CNR and B1+ signal amplification, suggesting that fabrication of complete 3D printed LC circuits for use as markers on catheters is feasible and can exhibit good tracking characteristics.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords