Integrated Parallel Reception, Excitation, and Shimming (iPRES) Breast Coil Array for Simultaneous MR Image Acquisition and Localized B0 Shimming

Yixin Ma1,2, Dean Darnell1, Huimin Zhang1,3, Fraser J Robb4, Allen Song1, and Trong-Kha Truong1,2,3

1Brain Imaging and Analysis Center, Duke University, Durham, NC, United States, 2Medical Physics Graduate Program, Duke University, Durham, NC, United States, 3Medical Physics Graduate Program, Duke Kunshan University, Kunshan, China, 4GE Healthcare, Aurora, OH, United States

In breast imaging, the image quality is severely degraded by susceptibility-induced B0 inhomogeneities. Here, we apply the novel integrated parallel reception, excitation, and shimming (iPRES) technology to enable simultaneous image acquisition and localized B0 shimming of the breasts with the same coil array. Proof-of-concept in vivo experiments with only four iPRES(1) breast coil elements already show up to 43.3% reduction in B0 root-mean-square error with no SNR compromise. Simulations show that the shimming can be further improved by implementing more complex iPRES(N) geometries with N smaller shim loops per RF coil element.