Meeting Banner
Abstract #1465

Estimated Measurement Uncertainty (EMU) in Calorimetrically-Determined Whole Body SAR Values for Medical Device Evaluation Using Benchtop Radiofrequency Exposure Systems

Krzysztof Wawrzyn1, Jack Hendriks1, William B. Handler1, and Blaine A. Chronik1

1The xMR Labs, Department of Physics and Astronomy, Western University, London, ON, Canada

The in vitro assessment of true radiofrequency whole body averaged specific absorption rate (WB-SAR) is described in the technical specification standard of ASTM F2182-11a, by direct measure of RF-induced heating within a standardized phantom centered inside the RF birdcage coil. F2182-11a does not address uncertainty assessment of the heating experiment. In this study, we present our measured values for short-term measurement repeatability and long-term measurement reproducibility. These measurements support the conclusion that RF-induced WB-SAR measurements made with bench-top RF exposure systems can be made with a total estimated measurement uncertainty of approximately 7% (k=1).

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords