Development of a radiolucent 64-channel on-body receive array to enhance image quality of the MR-linac

Stefan E. Zijlema1, Luca van Dijk1, Sara L. Hackett1, Jan J.W. Lagendijk1, Rob H.N. Tijssen1, and Cornelis A.T. van den Berg1

1Radiotherapy, UMC Utrecht, Utrecht, Netherlands

To improve the spatiotemporal resolution of 3D imaging on the MR-linac, we are developing a new radiolucent 64-channel receive array, which can be placed directly onto the patient during treatments. Coil prototypes caused no significant dosimetric changes. Measurements with 4-channel prototypes showed that overlapping coil loops lead to the highest potential imaging performance. Imaging comparisons with the current MR-linac array showed that the signal-to-noise ratio is improved.