Meeting Banner
Abstract #4183

The impact of pad thickness on RF burn prevention: is a 1mm pad sufficient?

Michael Steckner1 and Xin Chen1

1Toshiba Medical Research Institute USA, Mayfield Village, OH, United States

RF burns are typically categorized according to appearance and suspected mechanism. Skin/skin contact burns are postulated to be caused by significant current flows through a restricted contact area. While not definitively proven, patients are instructed to not clasp hands and pads should be routinely used to minimize skin/skin contact. The effectiveness of a thin pad is investigated with EM modeling. A reported calf burn (peak 10g SAR:whole body SAR ratio ~86) may be avoidable with a 1 mm air gap (SAR hotspot ratio ~2). Understanding that thin pads are sufficient may permit greater flexibility in possible pad solutions.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords