Using field monitoring to map and correct image distortions due to gradient nonlinearities

Ying-Hua Chu1,2, Yi-Cheng Hsu1,2, Fa-Hsuan Lin2, and Maxim Zaitsev1

1Dept. of Radiology, Medical Physics, Medical Center University of Freiburg, Faculty of Medicine, Freiburg, Germany, 2Institute of Biomedical Engineering, National Taiwan University, Taipei, Taiwan

We demonstrate how a field probe array can be used to measure gradient nonlinearities. Brain imaging results show that the distortion was most prominent at image voxels away from the iso-center. Compared to MRI vendor’s solution, our measurements and reconstruction could further reduce the image distortion by 35%.