Two prototypes of single-loop endoluminal receiver coils integrating serial and parallel active decoupling circuits using controllable MEMS switches were built and compared to a reference coil with PIN-diode circuit. Parallel MEMS coil had a significantly higher quality factor compared to the serial MEMS coil and the reference coil. The switching delays of both MEMS coils were about few microseconds (< 8µs). The characterization on 3.0T MRI system showed comparable signal intensity distribution for the different coil configurations.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords