The comparison of two ultra-short echo time methods for T2* mapping in Achilles tendon and enthesis

Vladimir Juras1,2, Peter Latta3, Martin Kojan3, Zenon Starcuk4, Xeni Deligianni5, Oliver Bieri5, and Siegfried Trattnig1

1High Field MR Centre, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Vienna, Austria, 2Institute of Measurement Science, Slovak Academy of Sciences, Bratislava 4, Slovakia, 3Central European Institute of Technology, Masaryk University, Brno, Czech Republic, 4Institute of Scientific Instruments, Czech Academy of Sciences, Brno, Czech Republic, 5Division of Radiological Physics, Department of Radiology, University of Basel Hospital, Basel, Switzerland

We compared the performance of two MR techniques, ultra-short TE (UTE) and variable-echo time (vTE), to estimate mono- and bi-exponential T2* in Achilles tendon and enthesis at 3T. In axial slice orientation, the mean T2* values for UTE were (mono, short, long) 4.1 ± 1.1 ms, 0.9 ± 0.3 ms and 19.6 ± 7 ms, for vTE we found the values 10.5 ± 3.9 ms, 1.4 ± 0.4 ms and 25 ± 3 ms. In sagittal slice orientation, the mean T2* values for UTE were (mono, short, long) 2.8 ± 0.5 ms, 0.4 ± 0.1 ms and 10.5 ± 1.7 ms, for vTE we found the values 5.3 ± 1.5 ms, 1.6 ± 0.3 ms and 22.3 ± 5.6 ms. T2* values calculated with vTE are overestimated in comparison to UTE, however, the clinical value of vTE might be higher, since it provides more slices with reduced blurring and lower scan times.