Meeting Banner
Abstract #1430

Use of passive B1+ shimming via dielectric pads for uniformity improvements in 7T clinical knee imaging

Andrew J Fagan1, Kimberly K Amrami1, Matthew A Frick1, Benjamin M Howe1, Peter Kollasch2, Venkata Chebrolu2, and Joel P Felmlee1

1Radiology, Mayo Clinic, Rochester, MN, United States, 2Siemens Healthineers, Rochester, MN, United States

Passive B1+ shimming was performed using custom dielectric pads to improve image uniformity for clinical knee imaging at 7T. Image non-uniformity is problematic for diagnosing knee pathologies, and post-processing solutions are limited at 7T due to the lack of a uniform body transmit coil. A range of pads made from high permittivity perovskite materials were fabricated and tested on 6 subjects. An optimal configuration was determined, comprising the pad dimensions and composition, commensurate with producing optimal image uniformity while fitting all subjects scanned. Dielectric pads represent an easy-to-use solution for improving clinical image uniformity at 7T.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords