31P-MR spectroscopy measures cardiac energetics in vivo directly by means of ATP and PCr, but is limited due to low SNR due to the low MR sensitivity of the 31P nucleus and limitation by the achievable B1+. Comprehensive RF coil design considerations and planning help fully exploit the SNR gain by increase of B0. In this simulation study we compare 28 different 31P 7T RF coil array designs with respect to their transmit field performance and obtain a 3-channel array as the best variant.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords