RF-induced heating pattern of a partially immersed conducting implant wire at 7T – its dependence on wire length

Pallab K Bhattacharyya1,2, Bhumi Bhusal3, Tanvir Baig3, Mark J Lowe1, Michael Martens3, and Stephen E Jones1

1Imaging Institute, Cleveland Clinic, Cleveland, OH, United States, 2Radiology, Cleveland Clinic Lerner College of Medicine, Cleveland, OH, United States, 3Physics, Case Western Reserve University, Cleveland, OH, United States

Radiofrequency induced heating at 7T of an insulated wire (exposed at tip) partially immersed in ASTM gel phantom to evaluate safety of scanning partially implanted guidewires, implants like stereo electroencephalography (SEEG) electrodes at different configurations, and conducting wires at that field strength. A transmit-receive coil was used for this study. Heating at certain resonance lengths separated by half a wavelength in air was observed. The heating at resonance lengths were 3-4 times less than that reported earlier at 3T at same specific absorption rate, and under similar configurations and settings.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords