Close-fitting the coil elements along the head and neck is a typical design of higher density phased-array head neck coils in order to provide higher SNR and acceleration. But this limits patient comfort and excludes larger patients. In this work we present a 64-channel head neck coil which can image large patients for neurovascular/fMRI at 3T. The coil test results show superior SNR in imaging compared to both a clinical 32-channel head array coil and 16-channel head neck array coil respectively.