Meeting Banner
Abstract #1587

Robust detection and potential application in orientation tracking of a small-amplitude field oscillation independent of location in the scanner bore and parallel to the static magnetic field.

Adam van Niekerk1, Andre van der Kouwe2,3, and Ernesta Meintjes1,4

1Human Biology, University of Cape Town, Cape Town, South Africa, 2Athinoula A. Martinos Center, Massachusetts General Hospital, Charlestown, MA, United States, 3Radiology, Harvard Medical School, Boston, MA, United States, 4Cape Universities Body Imaging Centre (CUBIC), University of Cape Town, Cape Town, South Africa

We explore the possible uses of a 40-kHz switching signal that is proportional to the gradient slew rate. The signal was identified by probing the rate of change of the magnetic flux density in the MRI scanner using a three-dimensional pickup coil. The signal is independent of both the orientation of the pickup coil and location in the scanner bore, and is directed parallel to the static magnetic field. This indicates that it is not caused by current in the gradient coils. These properties make this signal a useful vector reference that could be used for orientation tracking or time frame synchronisation.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords