Correction of B0 Inhomogeneities-Induced Artifacts in EPI for Breast Imaging

Ana Rodríguez-Soto1, Helen Park2, Dominic Holland3, Hauke Bartsch1, Haydee Ojeda-Fournier1, Anders Dale1, and Rebecca Rakow-Penner1

1Radiology, University of California, San Diego, La Jolla, CA, United States, 2School of Medicine, University of California, San Diego, La Jolla, CA, United States, 3Neuroscience, University of California, San Diego, La Jolla, CA, United States

Echo-planar imaging (EPI) is prone to B0 inhomogeneities that arise from susceptiblity effects resulting in spatial distortions more pronounced in the phase-encoded (PE) direction. Reduction and correction of such distortion artifacts are performed in both acquisition and post-processing stages. The present work evaluates the combined effect of a field-of-view (FOV) optimized and constrained undistorted single-shot (FOCUS) EPI and a distortion correction approach to that exploits the symmetry of the artifact on EPI images acquired with opposite PE polarities. After distortion correction good agreement was found between corrected EPI datasets and anatomical images.