Meeting Banner
Abstract #0705

Does Compliance with PIRADSv2 Technical Requirements Guarantee Image Quality or Adequacy in Prostate mpMRI Reads?

Jonathan J. Sackett1, Joanna Shih1, Sarah Reese1, Jeffrey R. Brender1,2, Stephanie Harmon1, Tristan Barrett1, Mehmet Coskun1, Manuel Madariaga 1, Jamie Marko 1, Yan Mee Law 1, Evrim Turkbey 1, Sherif Mehralivand 1, Thomas Sanford 1, Nathan Lay 1, Peter A. Pinto 1, Bradford J. Wood2, Peter L. Choyke2, Murali C. Krishna2, and Baris I Turkbey2
1Molecular Imaging Program, National Cancer Institute, National Institutes of Health, Bethesda, MD, MD, United States, 2National Cancer Institute, National Institutes of Health, Bethesda, MD, MD, United States

High quality MRIs are needed for prostate cancer screening and accurate targeting in MRI guided biopsies. Based on blind image quality assessment by image readers, compliance with the PIRADSv2 Minimum Technical Standards was determined to be neither necessary nor sufficient in ensuring quality in prostate T2 and DWI images.