Highly accelerated prostate water/fat 3D MRI with compressed sensing and SENSE, compared to the standard of care

Ya nan Wu1, Mei yu Sun1, Ai lian Liu1, Jia zheng Wang2, Yi shi Wang2, Cheng yan Wang1, Li hua Chen1, Qing wei Song1, Ren wang Pu1, and Bing bing Gao1
1The First Affiliated Hospital of Dalian Medical University, Da lian, China, 2Philips Healthcare,China, Bei jing, China

Compressed sensing(CS)technology is currently used less in prostate diseases. This study aims to explore the feasibility of CS and SENSE technology with different acceleration factors for prostate 3D magnetic resonance imaging(MRI). The preliminary exploration concluded that the SNR and CNR with CS 5 in prostate 3D mDIXON imaging are similar to the SENSE 2, and the imaging time is reduced by 28%, which effectively improves the MR scanning efficiency.