Explore performance of 3D MRCP with different compressed-sensing acceleration factors

Renwang Pu1, Jiazheng Wang2, Liangjie Lin2, Jingjun Wu1, Qingwei Song1, Nan Zhang1, and Ailian Liu1
1Department of Radiology, the First Affiliated Hospital of Dalian Medical University, Dalian, dalian, China, 2Philips Healthcare, Beijing, China, beijing, China

The present study aims to explore feasibility of three-dimensional(3D)breath-hold(BH)magnetic resonance cholangiopancreatography (MRCP) through acceleration of the combination of compressed sensing and sensitivity encoding(CS-SENSE),and choose an optimal acceleration factor. 3D-MRCP scans on healthy volunteers were carried out with different acceleration factors (conventional SENSE factor of 4, and CS-SENSE factors of18, 24, and 32).The CS-SENSE acceleration factor of 24 was recommended because of the favorable image quality and reasonable scan duration (15s).