Low field scanners present several advantages for interventional applications, however no standard method has yet arisen from temperature mapping. In this study we investigated the relationship between T1 and temperature at 0.1 T. MR images of four water phantoms, characterised by different T1, were acquired at different temperatures using a multi-slice FLASH sequence based on Look-Locker. Pixel-wise estimation of T1 relaxation through data fitting showed a growing trend with temperature. In light of the large number of time points, compromising SNR by a factor of 2 produced similar results for an almost 4 times faster acquisition.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords