Meeting Banner
Abstract #1392

Bailey splitter/combiner for RF shimming/Rx coil combination

Yue Zhu1,2, John C Gore1,2,3,4, and Xinqiang Yan1,2
1Vanderbilt University Institute of Imaging Science, Nashville, TN, United States, 2Department of Radiology and Radiological Sciences, Vanderbilt University, Nashville, TN, United States, 3Biomedical Engineering, Vanderbilt University, Nashville, TN, United States, 4Electrical and Computer Engineering, Vanderbilt University, Nashville, TN, United States

We designed a continuously adjustable Bailey splitter/combiner with a sliding mechanism. This device has a good input matching of < -18dB and a low insertion loss of about -0.5dB. The sliding mechanism allowed easy handling and precise adjustments. The resistor free design eliminated the potential safety concern from the previous Ratio Adjustable Power Splitter circuits. We obtained good agreements between predicted and measured transmit field maps with the Bailey splitter/combiner.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords