Meeting Banner
Abstract #3089

Direct Comparison of Gradient Modulation Transfer Functions and Acoustic Noise Spectra of the same MRI at High- (3T) and Lower-FieldĀ  (0.75T)

Hannes Dillinger1, Sebastian Kozerke1, and Christian Guenthner1
1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland

In this work, gradient fidelity is compared between high-field (3T) and lower-field (0.75T) configurations on the same MRI system by measuring the Gradient Modulation Transfer Function (GMTF) and audio noise spectra. Results demonstrate that at lower field mechanical resonances are reduced, leading to improved gradient fidelity and reduction of audio noise levels of up to 70%. Furthermore, it is shown that spectral peaks in the audio spectrum can be related to resonances in the GMTF, which suggests mechanical resonances can be identified using the audio spectrum.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords