Fast Multi-Parametric Mapping with an Extended Spiral and Off-Resonance Readout Correction

Willem van Valenberg1, Gyula Kotek1, Mika W. Vogel2, Juan A. Hernandez-Tamames1, and Dirk H. J. Poot1
1Radiology and Nuclear Medicine, Erasmus MC, Rotterdam, Netherlands, 2GE Healthcare, Hoevelaken, Netherlands

Synopsis

Single-shot quantitative methods with spiral sampling can increase the readout duration to reduce undersampling artifacts in the images and their propagation to the parameter maps. However, a longer readout increases blurring due to the off-resonance field if not accounted for in the reconstruction. We propose an efficient reconstruction that 1. corrects for blurring due to the off-resonance field, and 2. reduces undersampling artifacts by constraining the signal variation in each voxel to a predetermined subspace. The reconstruction method is shown to enable accurate multi-parametric mapping of a single-shot MP-b-nSSFP acquisition with submillimeter resolution in 1.8s.