Meeting Banner
Abstract #2633

Assessment of Femur Metallic Implant Safety at 3T

Amira Trabelsi1,2,3, Megdouda Benamara4, Pierre Jomin1, Stefan Enoch1, Marc Dubois4, Martine Pithioux3, David Bendahan2, and Redha Abdeddaim1
1Aix Marseille Univ, CNRS, Centrale Marseille, Institut Fresnel, Marseille, France, 2Aix Marseille Univ, CNRS, CRMBM, Marseille, France, 3Aix Marseille Univ, CNRS, ISM, Marseille, France, 4Multiwave Imaging, Marseille, France

Synopsis

We assessed electromagnetic behavior of metallic implant used for distal femur fracture in typical MRI situation. |B1+| field, SAR and temperature variations were computed at 3T using a surface coil on human model for different positions of the implant relative to the coil. To validate the simulation, we measured the E-field and compared it to simulated E-field. |B1+| field maps showed an interesting augmentation near the implant. Both global SAR and local SAR levels proved that it is possible to safely image bone repair. However, temperature elevation near the tip of the implant was important and is to be considerate.