Meeting Banner
Abstract #5029

A Proposed Method for Retuning an RF Coil by Changing the Shield Radius for Multinuclear Imaging

Kieffer Davieau1, Matthew Fox1,2, Blaine Chronik1,3, and Alexei Ouriadov3,4
1Medical Biophysics, Western University, London, ON, Canada, 2Lawson Health Research Institute, London, ON, Canada, 3Physics and Astronomy, Western University, London, ON, Canada, 4School of Biomedical Engineering, Western University, London, ON, Canada

Synopsis

This study investigates the feasibility of changing the shield radius of an RF coil to shift the resonant frequency from 1H to 19F for multinuclear imaging. The method proposed in this investigation was verified by changing the resonant frequency of a 129Xe RF Coil a total of 1.18 MHz by changing the shield radius, allowing it to be used in a 2.9T Siemens MRI and a 3.0T GE MRI. It was found that altering the shield radius had similar effects on the resonant frequency as using a variable tuning capacitor, allowing for one coil to be compatible with multiple scanners.

This abstract and the presentation materials are available to members only; a login is required.

Join Here