Keywords: Low-Field MRI, Low-Field MRI, back imagingA compact and lightweight single-sided Inward-outward (IO)-ring permanent magnet array is proposed for back imaging. Its size is 1396×528×1080mm3 and it weighs 924 kg. Within a field-of-view (FoV) of 300×80×20mm3, it has an axial magnetic field with an average strength of 129.8mT, a built-in gradient (averaged at 158mT/m) that corresponds to a radiofrequency (RF) bandwidth of 9.7% and a good linearity with a R2-coefficient of 0.952. It has a compact 5-Gauss zone of 400×320×460mm3.