Meeting Banner
Abstract #0598

Simulated radiation patterns of MRI without a shielded room from 0.5 to 7 Tesla

Ehsan Kazemivalipour1,2, Bastien Guerin1,2, and Lawrence L. Wald1,2,3
1A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 2Harvard Medical school, Boston, MA, United States, 3Harvard-MIT Division of Health Sciences Technology, Cambridge, MA, United States

Synopsis

Keywords: Safety, SafetyFar-field electromagnetic radiation patterns and levels were simulated on the 10m radius regulatory sphere for conventional MRI scanners at 0.5T, 1.5T, 3T, and 7T operated without an RF shielded room. The levels and patterns were strongly affected by the symmetry of the load. With a body load, the peak E-fields on a 10m radius surface rose with roughly the square of the frequency and far exceeded regulatory limits even for 0.5T. With the body in the bore, the radiated patterns also take on a highly asymmetric pattern not present for more symmetric loads.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords