7 Tesla MRI of the knee joint after anterior cruciate ligament graft reconstruction: safety discussion and image quality study

Oliver Kraff1, Jens M Theysohn2, Jana Theisejans3, and Harald H Quick1,4
1Erwin L Hahn Institute for MRI, University Duisburg-Essen, Essen, Germany, 2Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany, 3General Psychology: Cognition and Center for Behavioral Addiction Research (CeBAR), University Duisburg-Essen, Duisburg, Germany, 4High-Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany

Synopsis

Keywords: Safety, ArtifactsThis study presents the MR safety discussion for imaging subjects with metallic fixation buttons after reconstruction of the anterior cruciate ligament, which have not been labeled MR conditional at 7T by the implant vendor. In addition, image quality and artifacts are evaluated in a knee imaging protocol consisting of both gradient- and spin echo sequences. Two imaging cases are presented: one with 7T imaging before and after surgery, and another case with comparative imaging between 1.5T and 7T. Artifact sizes from metallic fixation buttons at the femur did not impair the image quality and diagnostic evaluation of the knee joint.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords