Meeting Banner
Abstract #0613

All-inclusive Safety Models Limit Imaging Performance at 7T

Emre Kopanoglu1
1CUBRIC, School of Psychology, Cardiff University, Cardiff, United Kingdom

Synopsis

Keywords: Safety, High-Field MRI, RF Pulse Design & Fields; Parallel Transmit & MultibandSafety models on scanners are patient-position unaware and may include multiple patient positions to ensure safety. This may lead to overestimation of the specific absorption rate (SAR) and limit scanning performance. This study investigates the effect of an all-inclusive safety model on SAR estimation for parallel-transmit and quadrature-excitation at 7T. Results show that more than 4-fold SAR overestimation can be commonly observed. RF shimming suffered the most, with 11-fold overestimation at the worst-case and more than 4-fold overestimation in 37% of cases. RF shimming also offered the lowest peak local SAR and may be unnecessarily penalized by overconservative safety models.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords