Keywords: Shims, ShimsIn this work, we propose a reverse RF/B0 shimming coil design for ultrahigh field MRI without using any large bridge choke inductors. Our simulation and experimental results reveal that this design does not impair the RF performance, but would able to reduce the DC resistance/inductance/heating and also save precious space in RF/B0 shimming coils.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords