‘Reverse’ RF/B0 shimming coils

Xinqiang Yan1,2, Shuyang Chai1,2, Ming Lu1,2, and John C Gore1,2
1Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, TN, United States, 2Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, United States

Synopsis

Keywords: Shims, ShimsIn this work, we propose a reverse RF/B0 shimming coil design for ultrahigh field MRI without using any large bridge choke inductors. Our simulation and experimental results reveal that this design does not impair the RF performance, but would able to reduce the DC resistance/inductance/heating and also save precious space in RF/B0 shimming coils.