Test-Retest Repeatability and Cross-Platform Reproducibility of Optimized 2D MRE and 3D MRE Liver Protocols with Automated Analysis

Bogdan Dzyubak1, Kay M Pepin1, Jun Chen1, Yuan Le1, Kyle Kalutkiewicz2, Roger Grimm1, Jeremy A Heilman1, Scott Kruse1, Jennifer Kugel1, Meng Yin1, Kevin J Glaser1, and Ehman L. Richard1
1Radiology, Mayo Clinic, Rochester, MN, United States, 2Engineering, Resoundant, Rochester, MN, United States

Synopsis

Keywords: Liver, ElastographyMagnetic Resonance Elastography is a technique used to effectively stage liver fibrosis by measuring tissue stiffness. This work optimized several acquisition and processing parameters, and performed a comprehensive repeatability/reproducibility validation across the three types of MRE (GRE, EPI, 3DEPI), 1.5T and 3T field strengths, and the three major manufacturers (GE, Siemens, Philips). When using the optimized methods and automated analysis, the repeatability coefficients were 11.8% for GRE, 10.8% for EPI, and 7.7% for 3DEPI, all superior to QIBA’s 19% benchmark. Agreement between manufacturers was within 2%.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords