Keywords: Data Analysis, SpectroscopyThe effect of baseline flexibility LCModel parameter DKNTMN (0.15, 0.3, 0.6 and 1) on a 7T short-TE STEAM MRS measurement of GABA, glutamate and excitatory-inhibitory ratio (EIR) was investigated using both simulated and measured macromolecular basis sets. Mean (SD) of GABA/tCr was highest, 0.23 (0.02), using simulated MM basis. Using measured MM basis, the ratios decreased from 0.18 (0.04) to 0.12 (0.03) by the increase of DKNTMN values. The GABA/tCr ratio and EIR of a former multi-center study was 0.19 and 8.2 after T2 decay correction. Analysis using DKNTMN of 0.3 conformed best and considered to be the choice.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords