In-bore Preamplifier Nonlinearity Calibration

Chris Vassos1, John Pauly1, Fraser Robb2, Shreyas Vasanawala3, and Greig Scott1
1Electrical Engineering, Stanford University, Stanford, CA, United States, 2GE Healthcare, Aurora, OH, United States, 3Radiology, Stanford University, Stanford, CA, United States

Synopsis

Keywords: Hybrid & Novel Systems Technology, System Imperfections: Measurement & CorrectionPresent MRI preamplifiers are very linear but use 10s of mA bias current and over 150mW power. For wireless arrays, low power alternatives based upon SiGe transistors are needed. These achieve the needed gain and noise figure, with order of magnitude power reduction, but at the cost of increased nonlinearity. Using pilot tone technology, we present a calibration system for preamplifier nonlinearity correction targeted to low power SiGe preamplifiers. This system integrated a phase-locked SDR into a form of prescan to demonstrate that in-bore preamplifier linearity calibration and correction are possible.