Considerations for Accurate Active Detuning Measurements in Densely Parallel Array Coils

Manisha Shrestha1, Mirsad Mahmutovic1, Sam-Luca JD Hansen1, Jason P Stockmann2, and Boris Keil1
1Institute of Medical Physics and Radiation Protection, TH Mittelhessen University of Applied Sciences, Giessen, Germany, 2A.A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Harvard Medical School, Boston, MA, United States

Synopsis

Keywords: RF Arrays & Systems, High-Field MRI, detuning

Active detuning of receive array coil during transmit prevents large currents being induced in the coil which could damage sensitive electronic components and create SAR hotspots. The standard method of adjusting active detuning using a double probe shows an offset in the resonant frequency set at bench when tested in the presence of a scanner. Setting a high impedance at the input of loop using S11 measurement exactly at Larmor frequency has proven to be a better method to detune loops actively.