Determination of the Optimum Permittivity of High-Permittivity Pads when Dipoles are Used as Transmit Coils for the Chest

Giuseppe Carluccio1,2 and Christopher Michael Collins1,2
1Radiology, NYU Grossman School of Medicine, New York, NY, United States, 2Center for Advanced Imaging Innovation and Research (CAI2R), New York University, New York, NY, United States

Synopsis

Keywords: RF Arrays & Systems, RF Arrays & Systems, High-Permittivity MaterialsHigh-Permittivity Materials are often used to manipulate the radiofrequency field B1. Optimization of the permittivity depends on the geometry of the pad, of the subject and of the coil. In this work we determine the permittivity of two pads positioned around the torso, having thickness 1.5 cm and 3 cm, when a dipole array is used to provide the highest transmit efficiency in the heart. The pads provide an improvement of the transmit efficiency of 9.5% and 7.6% for the 1.5 cm and 3cm case respectively, while providing a SAR reduction of 20.2% and 12.5%.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords