Evaluation of MRI performance and PET compatibility of a 32 channel coil for the Siemens Biograph mMR PET/MRI scanner

Harris Bidaman1, Katie Dinelle1, Reggie Taylor2,3, and Lauri Tuominen2,4
1Brain Imaging Centre, Royal Ottawa Mental Health Centre, Ottawa, ON, Canada, 2University of Ottawa Institute of Mental Health Research at The Royal, Ottawa, ON, Canada, 3Department of Physics, Carleton University, Ottawa, ON, Canada, 4Department of Psychiatry, University of Ottawa, Ottawa, ON, Canada

Synopsis

Keywords: PET/MR, Brain, CoilWe evaluate MRI and PET performance of the novel Siemens 32 channel head coil for the Siemens Biograph mMR. Results are compared to the system standard 12 channel coil. MRI performance (SNR) was improved with the 32 channel coil. PET image uniformity was not impacted. ROI analysis of the PET data indicated a 5% underestimation of activity concentration in the 32 channel coil, however this difference was mitigated by the use of an image derived reference region. Any small differences in PET performance introduced by the 32 channel coil are far outweighed by the superior MRI performance of this coil.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords