Assessment and mitigation of EMI from in-room equipment in the setting of interventional 0.55T MRI

Bilal Tasdelen1, Ecrin Yagiz1, Rajiv Ramasawmy2, Ahsan Javed2, Dogangun Uzun2, Dursun Korel Yildirim2, Adrienne Campbell-Washburn2, and Krishna S Nayak1
1Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, CA, United States, 2Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD, United States

Synopsis

Keywords: System Imperfections: Measurement & Correction, System Imperfections: Measurement & Correction, EMI, interference, pilot toneFive devices that are relevant in interventional setting and a Pilot-Tone generator are assessed in terms of their electromagnetic interference with 0.55T MRI. Two of the devices (ergometer and flat-panel television) generated significant interference, however acceptable image quality was achieved when using the EDITER interference cancellation technique. EDITER also enabled the use of higher Pilot-Tone amplitudes and use of cheap, general-purpose signal generator for Pilot-Tone generation.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords