Keywords: Hybrid & Novel Systems Technology, Prostate, Eddy-current, GIRFDiffusion-weighted imaging requires strong sensitizing gradients, resulting in eddy currents that create image distortions and hinder the reliability of measured diffusion parameters. In this study, we characterize and and compensate for eddy currents on a compact inside-out nonlinear gradient using a 2D phase encoded FLASH sequence. Time series of the effective waveform was calculated from PCA of the field maps, which was then used to calculate a gradient impulse response function (GIRF). The GIRF was validated by comparing the output waveform of a simple vs pre-emphasized trapezoid. Results showed clear improvement of the pre-emphasized waveform.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords