Keywords: RF Arrays & Systems, Body
The purpose of this study is to compare two different coil array designs for prostate imaging. We compared an in-house developed 8chpTx/Rx micro stripline element coil array with meanders to a recently introduced, commercially available 8chpTx/Rx coil array based on fractionated dipoles for 1H prostate imaging at 7T. Both coil arrays showed comparable results in the phantom and in vivo examinations. The commercially available coil array performs slightly better in the region of the prostate and is well suited for future prostate imaging at 7T.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords