Keywords: Parallel Imaging, Vessels, compressed sensitivity encodingThe clinical application of phase-contrast magnetic resonance venogram (PC-MRV) is limited by the long scan time. This study aims to accelerate the acquisition of PC-MRV by the compressed sensitivity encoding (CS-SENSE) technique and find the optimal acceleration factor. Results show that the image quality of fast PC-MRV based on CS-SENSE was significantly higher than that based on conventional sensitivity encoding (SENSE) technology, and the CS-SENSE factor 9 was recommended for mild to moderate patients, and CS AF=12 for critical patients.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords