Keywords: Relaxometry, Phantoms, TemperaturePhantom vials of varying T2 were calibrated as a function of temperature, where the temperature was measured via multi-echo gradient echo imaging of ethylene glycol in a phantom vial. The temperature calibrations were used to adjust T2 to a reference temperature, achieving T2 deviations to within [-3.2, 2.0] ms [-3.1%, 2.9%] on Siemens 3T (16.7ºC – 25.1ºC) and [-3.5, 1.4] ms [-3.4%, 0.8%] on GE 1.5T (16.5ºC – 23.7ºC). This demonstrates that corrections for T2 due to temperature can be performed while measuring temperature from gradient-echo imaging of ethylene glycol.