Applicability of Magnetic Resonance Fingerprinting for Proton Resonance Frequency Shift based Temperature Monitoring of Microwave Ablation

Moritz Gutt1, Josef Joaquin Löning Caballero2, Dominik Horstmann2, Frank Wacker2, Bennet Hensen2, and Marcel Gutberlet2
1Computational Health Informatics, Leibniz University Hannover, Hannover, Germany, 2Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany

Synopsis

Keywords: Thermometry, MR FingerprintingMRI offers the unique ability of measuring the temperature inside of the patient during minimally invasive tumor ablation. In this proof-of-concept approach MR Fingerprinting was used in order to get an accurate thermometry with potential for real-time application. The method was validated by performing a microwave ablation in a bioprotein phantom. The temperature accuracy was 1.54 °C ± 0.85 °C and the dice score between the calculated and the ground truth denaturation zone was 90.02 %. The mean standard deviation of the non-heated area was 0.19 °C.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords