Meeting Banner
Abstract #4996

Optimization of EPI based PRF thermometry at 0.5 T

Chad T Harris1, Curtis N Wiens1, Diego F Martinez2, Andrew T Curtis1, and Blaine A Chronik2
1Research and Development, Synaptive Medical, Toronto, ON, Canada, 2The xMR Labs, Department of Physics and Astronomy, Western University, London, ON, Canada

Synopsis

Keywords: Data Acquisition, ThermometryEcho planar imaging (EPI) is well-suited for temperature mapping in the mid-field (0.5T) thanks to longer T2* values that can be leveraged without prohibitively slow update rates. In this work, we optimize EPI based PRF thermometry acquisitions for mid-field MRI. Digital simulations and in-vivo temperature stabilities were computed over echo times ranging from 50ms – 100ms. The optimal mean temperature uncertainty over the entire acquisition volume was measured to be approximately 1⁰C demonstrating that EPI based thermometry provides clinically viable sensitivity at 0.5T with sufficient image resolution and extremely fast update rates (1.3-2 s).

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords