Keywords: Non-Array RF Coils, Antennas & Waveguides, Non-Array RF Coils, Antennas & WaveguidesMulti-frequency coils are typically known to have substantial losses as compared to their single-tuned counterparts. Here we investigate sensitivity tradeoffs of triple-tuned (1H-23Na-31P) trap designs on receive coils in a tightly controlled experiment to isolate the sensitivity effects of the traps. Triple-tuned LCC traps were shown to have significant improvement in sensitivity for 23Na and 1H as compared to an LC trap design. The increase came at the expense of 31P sensitivity, however continued optimization of the trap design will likely further improve the final experimental SNR.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords