Remarkably enhanced B1 and SNR with uHDC ceramics integrated with novel RF transreceiver array for 2H, 17O, and 1H imaging of human brain at 10.5T

Soo Han Soon1,2, Matt Waks1, Xin Li1, Hannes M. Wiesner1, Xiao-Hong Zhu1, and Wei Chen1,2
1Center of Magnetic Resonance Research (CMRR), Department of Radiology, University of Minnesota, Minneapolis, MN, United States, 2Department of Biomedical Engineering, University of Minnesota, Minneapolis, MN, United States

Synopsis

Keywords: RF Arrays & Systems, Parallel Transmit & Multiband, High Permittivity Material (HPM), ultrahigh Dielectric Constant (uHDC) Material, MRSI, UHF, Ultrahigh field, Broadband RF Coil

Motivation: 2H and 17O MRSI are useful to study brain energy metabolism, however, have low imaging sensitivity.

Goal(s): To develop a RF coil engineering solution offering superior performance for human brain 1H MRI, 2H and 17O MRSI at UHF.

Approach: We constructed a novel RF transreceiver array coil, which can operate at 1H, 2H and 17O resonant frequencies at 10.5T. The integration of ultrahigh dielectric constant (uHDC) ceramics enhanced RF transmission field (B1+) and SNR.

Results: The new array coil functioned well for performing multinuclear imaging, and the integrated uHDC technology remarkably enhanced B1+ and SNR for performing 2H and 17O MRSI.

Impact: In this study, we introduce and demonstrate an advanced RF array coil integrated with the uHDC material enabling imaging of three important nuclei (1H, 2H and 17O) signals with superior performance aiming for human brain applications at UHF of 10.5T.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords